Fluke 754 Documenting Process Calibrator with HART Instrument
$9,250
Model : 754 Calibrator with HART Instrument
Description
Fluke 754 Documenting Process Calibrator with HART Instrument
Fluke 754 Documenting Process is a multi-function documentation calibrator that allows you to download software-generated procedures, lists, and instructions or upload data for printing, archiving, and analysis. The Model 754 also features a powerful built-in HART® interface that performs the same everyday tasks you would with a separate communicator.
Key Feature Fluke 754 Documenting Process
- Measure voltage, mA, RTD, thermocouple, frequency, and ohms to test sensors, transmitters, and other devices.
- Source/simulate voltage, mA, thermocouple, RTD, frequency, ohms, and pressure to calibrate transmitters
- Generating the transmitter during testing using a loop supply while simultaneously performing mA measurements
- Measure/source pressure using one of 29 Fluke 700Pxx Pressure Modules
- Create and execute as-found/as-left procedures to meet quality programs or regulations. Record and document results.
- Saves downloaded calibration procedures and results for a full week.
- Using many features such as auto-stage, custom units, user-entered values during testing, single-point and double-point switch testing, square root DP flow testing, programmable measurement delays, etc.
- Easy to use
- Three year warranty
- Bright white dual display. Read source parameters and measured parameters simultaneously.
- Multi language interface
- Rechargeable Li-Ion battery for 10 hours of continuous use. Includes gas gauge.
- Handles fast pulse RTD transmitters and PLCs, with pulses as short as 1 ms.
- Delivered with DPC/Track Sample software.
- Compatible with various Asset Management software packages.
The Model 754 HART capability
Is designed to do what you do on a daily basis with a separate communicator. In fact, it offers the communication capabilities of the 375 HART communicator.
- No external box or additional tools are required for HART daily calibration and maintenance.
- Offers fast HART communications.
- Supports popular models of HART transmitters, with support for more specific commands than any other HART field calibrator.
- Works with multiple master, burst modes, and multidrop configurations.
- Easy to update as additional devices are added and new versions of HART are launched.
- Interrogate to determine device type, manufacturer, model, tags.
- Reconfiguring dual sensor temperature transmitter mapping.
- Reads HART PV function and transmitter digital output while measuring analog mA output.
- Read and write configuration functions to perform field adjustments to PV span points, damping, and top-level configuration settings.
- Relabel smart transmitters by reading and writing HART tags.
Versatile HART Protocol support
Manufacturer | Pressure device | Temperature device | Coriolis device |
---|---|---|---|
ABB/Kent Taylor | 600T | 658T | |
ABB/Hartmann & Braun | Contrans P, AS 800 Series |
||
Endress & Hauser | CERABAR S, CERABAR M, DELTABAR S |
TMT 122 , TMT 182 , TMT 162 |
|
Foxboro Eckardt | TI/RTT20 | ||
Foxboro/Invensys | I/A Pressure | ||
Fuji | FCX FCXAZ |
FRC | |
Honeywell | ST3000 | STT25T, STT25H |
|
Micro Motion | 2000 2000 IS 9701 9712 9739 |
||
Moore Products | 344 | ||
Rosemount | 1151 2088 3001C 3051, 3051S |
3044C 644 3144 3244, 3144P |
|
Siemens | SITRANS P DS SITRANS P ES |
||
SMAR | LD301 | TT301 | |
Viatran | I/A Pressure | ||
Wika | UNIT TRANS | T32H | |
Yokogawa | SPELL | YTA 110, 310 and 320 |
Supported HART Operating Modes
- Point-to-Point operation, the most commonly used mode, connects the model 754 to a HART device in a 4-20 mA loop.
- Multi-Drop mode, multiple HART devices can be bused at once. The 754 looks for each one, identifies the address used, and lets you select the device for calibration and related operations.
- In burst mode, the HART device transfers data without waiting to be interrogated by the master unit. The Model 754 can take the transmitter out of burst mode during testing or calibration, then return it to burst mode.
Pressure Module Fluke 754 Documenting Process
Covers almost all pressure applications including gauge, differential, dual (mixed), absolute and vacuum pressures.
- Displays pressure readings in 10 different pressure units that you specify in the calibrator settings.
- The rugged urethane box protects the module from rough handling and harsh conditions.
- Features internal temperature compensation from 0º to 50º C for full accuracy performance.
- Accompanied by a calibration certificate traceable to NIST.
- Modules can be calibrated locally, helping to control costs.
A suite of 29 optional pressure modules provides pressure calibration and measurement capabilities. Twenty-eight modules are available, with accuracy specifications to 0.05%. Ranges start at 0 — 1 in H20 (0 — 0.25kPa) and go to 0 — 10,000 psi (0 — 70,000 kPa). Additional information about pressure modules is available on the Pressure Modules Home Page.
Automated Procedures Fluke 754 Documenting Process
Enables you to quickly set up automated calibration procedures for linear transmitters, DP flow transmitters, and one- and two-point limit switches. Simply select the appropriate source size and/or function and fill in the procedure template. The 750 Series does the rest. It performs the test, calculates the error, and quickly displays the final results, highlighting the tolerance points.
Adjustment Unit Fluke 754 Documenting Process
Allows you to map between units, such as mV to ºC or ºF. Allows you to use the Fluke 750 Series with millivolt output accessories such as the 80T-IR Temperature Probe, and to document tests using unsupported units such as parts per million (ppm) or revolutions per minute.
Differential Pressure Flow Instrument Calibration
Use the square root function routinely to calibrate DP flow devices directly.
Multifunction Additional Features
Calibrate temperature, pressure, voltage, current, resistance, and frequency. Because the device measures while it’s running, you can troubleshoot and calibrate everything with one powerful tool.
Powerful, easy to use Fluke 754 Documenting Process
An integrated, easy-to-follow menu display guides you through each task. Get up to speed in minutes, not days. Programmable calibration routines allow you to create and run as-is/as-left procedures to ensure fast, consistent calibrations.
Record and document results
To meet ISO-9000 or regulatory standards, the Fluke 754 records calibration results for you, so you don’t have to lug a pen and pad into the field. A USB interface lets you transfer results to a PC, saving you time transcribing manually when you get back to the shop.
Just one handful Fluke 754 Documenting Process
Small enough to fit easily in a tool bag and use in tight spaces. Powered by a rechargeable Li-ion battery pack.
Strong and reliable Fluke 754 Documenting Process
Rely on Fluke’s rugged design for superior accuracy and durability in rugged environments. The urethane housing anticipates rough handling in industrial environments.
Bright white display
Allows you to read the results in both light and dark conditions. The backlight has three (3) settings.
Soft keys Fluke 754 Documenting Process
One-touch access is available for advanced functions such as task lists, automated procedures, inspections, min/max, staging and ramping, and review memory.
Specifications Fluke 754 Documenting Process
Measurement Accuracy | |
Voltage DC | 1 Year 2 Years 100.000 mV 0.02%+0.005 mV 0.03%+0.005 mV 3.00000 V 0.02%+0.00005 V 0.03%+0.00005 V 30.0000 V 0.02%+0.0005 V 0.03%+0.0005 V 300.00 V 0.05%+0.05 V 0.07%+0.05 V |
Voltage AC | Range 40 to 500Hz Resolution 1 Year 2 Years 3.000 V 0.001 V 0.5%+0.002 V 1.0%+0.004 V 30.00 V 0.01 V 0.5%+0.02 V 1.0%+0.04 V 300.0 V 0.1 V 0.5%+0.2 V 1.0%+0.2 V |
Current DC | 1 Year 2 Years 30.000 mA 0.01% + 5 uA 0.015%+7 uA 110.00 mA 0.01% + 20 uA 0.015%+30 uA |
Resistance | 1 Year 2 Years 10.000 O 0.05% + 50 mO 0.07%+70 mO 100.00 O 0.05% + 50 mO 0.07%+70 mO 1.0000 kO 0.05% + 500 mO 0.07%+0.5 O 10.000 kO 0.1% + 10 O 0.15%+15 O |
Frequency | Resolution Accuracy (2 Years) 1.00 to 110.00 Hz 0.01 Hz 0.05 Hz 110.1 to 1100.0 Hz 0.1 Hz 0.5 Hz 1.101 to 11.000 kHz 0.001 kHz 0.005 kHz 11.01 to 50.00 kHz 0.01 kHz 0.05 kHz |
Source Accuracy | |
Voltage DC | 1 Year 2 Years 100.000 mV 0.01%+0.005 mV 0.015%+0.005 mV 1.00000 V 0.01%+0.00005 V 0.015%+0.0005 V 15.0000 V 0.01%+0.0005 V 0.015%+0.0005 V |
Current DC | 1 Year 2 Years 22.000 mA (Source) 0.01%+ 0.003 mA 0.02%+ 0.003 mA Current Sink (Simulate) 0.02% + 0.007 mA 0.04% + 0.007 mA |
Resistance | 1 Year 2 Years 10.000 O 0.01% + 10 mO 0.015% + 15 mO 100.00 O 0.01% + 20 mO 0.015% + 30 mO 1.0000 kO 0.02% + 0.2 O 0.03% + 0.3 O 10.000 kO 0.02% + 3 O 0.03% + 5 O |
Frequency | 2 Years 0.1 to 10.99 Hz 0.01 Hz 0.01 to 10.99 Hz 0.01 Hz 11.00 to 109.99 Hz 0.1 Hz 110.0 to 1099.9 Hz 0.1 Hz 1.100 to 21.999 kHz 0.002 kHz 22.000 to 50.000 kHz 0.005 kHz |
Technical Data | |
Data Log Functions | Measure functions: Voltage, current, resistance, frequency, temperature, pressure Reading rate: 1, 2, 5, 10, 20, 30, or 60 readings/minute Maximum record length: 8000 readings (7980 for 30 or 60 readings/minute) |
Ramp Functions | Source functions: Voltage, current, resistance, frequency, temperature Rate: 4 steps/second Trip detect: Continuity or voltage (continuity detection not available when sourcing current) |
Loop Power Function | Voltage: Selectable, 26 V Accuracy: 10%, 18 V minimum at 22 mA Maximum current: 25 mA, short circuit protected Maximum input voltage: 50 V DC |
Step Functions | Source Functions Voltage, current, resistance, frequency, temperature Manual Step Selectable step, change with arrow buttons Autostep Fully programmable for function, start delay, stepvalue, time per step, repeat |
Environmental Specifications | |
Operating Temperature | -10 °C to +50 °C |
Storage Temperature | -20 °C to +60 °C |
Dust/water resistance | Meets IP52, IEC 529 |
Operating Altitude | 3000 m above mean sea level (9842 ft) |
Mechanical & General Specifications | |
Size | 136 x 245 x 63 mm (5.4モ x 9.6 x 2.5) |
Weight | 1.2 kg (2.7 lb) |
Batteries | Internal Battery Pack Li-ion: 7.2V, 4400mAh, 30 Wh |
Battery Life | >8 hours typical |
Battery Replacement | Replace without opening calibrator; no tools required |
Side port connections | Pressure module connector USB Connector to interface to your PC Digital instrument (HART) connector Connection for optional battery charger/eliminator |
Data storage capacity | 1 week of calibration procedures results |
Temperature, Resistance Temperature Detectors | |||||||
Degrees or % of reading | |||||||
Type (a) | Range °C | Measure °C1 | Source current | Source °C | Allowable current | ||
1 year | 2 year | 1 year | 2 year | ||||
100 O Pt (385) | -200 to 100 100 to 800 |
0.07 °C 0.02 % + 0.05 °C |
0.14 °C 0.04 % + 0.10 °C |
1 mA | 0.05 °C 0.0125 % + 0.04 °C |
0.10 °C 0.025 % + 0.08 °C |
0.1 mA to 10 mA |
200 O Pt (385) | -200 to 100 100 to 630 |
0.07 °C 0.02 % + 0.05 °C |
0.14 °C 0.04 % + 0.10 °C |
500 µA | 0.06 °C 0.017 % + 0.05 °C |
0.12 °C 0.034 % + 0.10 °C |
0.1 mA to 1 mA |
500 O Pt (385) | -200 to 100 100 to 630 |
0.07 °C 0.02 % + 0.05 °C |
0.14 °C 0.04 % + 0.10 °C |
250 µA | 0.06 °C 0.017% + 0.05 °C |
0.12 °C 0.034 % + 0.10 °C |
0.1 mA to 1 mA |
1000 O Pt (385) | -200 to 100 100 to 630 |
0.07 °C 0.02 % + 0.05 °C |
0.14 °C 0.04% + 0.10 °C |
150 µA | 0.06 C 0.017 % + 0.05 °C |
0.12 C 0.034 % + 0.10 °C |
0.1 mA to 1 mA |
100 O Pt (3916) | -200 to 100 100 to 630 |
0.07 °C 0.02 % +0.05 °C |
0.14 °C 0.04 % +0.10 °C |
1 mA | 0.05 °C 0.0125 % + 0.04 °C |
0.10 °C 0.025 % + 0.08 °C |
0.1 mA to 10 mA |
100 O Pt (3926) | -200 to 100 100 to 630 |
0.08 °C 0.02 % +0.06 °C |
0.16 °C 0.04 % +0.12 °C |
1 mA | 0.05 °C 0.0125 % + 0.04 °C |
0.10 °C 0.025 % + 0.08 °C |
0.1 mA to 10 mA |
10 O Cu (427) | -100 to 260 | 0.2 °C | 0.4 °C | 3 mA | 0.2 °C | 0.4 °C | 0.1 mA to 10 mA |
120 O Ni (672) | -80 to 260 | 0.1 °C | 0.2 °C | 1 mA | 0.04 °C | 0.08 °C | 0.1 mA to 10 mA |
Package Include :
- Fluke 754 Documenting Process Calibrator
- (6) Fluke Flexible Test Leads
- (6) Fluke Test Pins
- (6) Fluke Alligator Clips
- (4) Fluke Hook Clips
- Hart Communication Cable
- Fluke Jumper Cable
- Strap For Unit
- AC Adapter/Charger
- USB PC Interface Cable
- Hard Copy User Manual
- User Manual On CD-ROM
- Carrying Case w/ Strap
Additional information
Lead Time | 5-7 days |
---|---|
Packaging | Box |
Reviews
There are no reviews yet.