Fluke ESA612 Electrical Safety Analyzer 230V
Fluke ESA612 combines the power of a ECG simulator and electrical safety analyzer in a single rugged portable device includes two electrical safety test load selections for global use, as well as one-button access to multiple applications.
The ESA612 Electrical Safety Analyzer represents the next generation in testers for biomedical professionals that perform field service on medical equipment throughout their facilities, in clinics, and anywhere onsite service is required. Portable, lightweight, and designed for operation in tight spaces, Fluke ESA612 offers the functionality of a simulator, multimeter and electrical-safety analyzer in a single test tool.
With selection of two test loads, this versatile product can be used worldwide to test to preventative maintenance electrical safety standards of choice: ANSI/AAMI ES1:1993 (NFPA-99), IEC62353 (VDE 751), and AN/NZS 3551.
The versatility of the multifaceted ESA612 is further expanded with optional automation software, which speeds and simplifies testing and provides high-end-analyzer productivity at software-level investment. Ansur-automated ESA612 standardizes test procedures, compares results to standards limits, and generates and stores reports for total digital data management.
The power of a ECG simulator and electrical safety analyzer in a single rugged portable device.
Fluke ESA612 Electrical Safety Analyzer represents over 20 years of Fluke Biomedical experience in test equipment for biomedical and clinical engineering professionals that perform field service on medical equipment throughout their facilities, in clinics, and anywhere onsite service is required. Portable, lightweight, and designed for operation in tight spaces, the ESA612 offers the functionality of a simulator and electrical safety tester in a single test tool. With the selection of two test loads, this versatile product can be used worldwide for compliance testing to preventative maintenance electrical safety standards of choice. The versatility of the multifaceted ESA612 Electrical Safety Analyzer is further expanded with optional automation software, which speeds and simplifies testing and provides high-end-analyzer productivity at software-level investment. Ansur-automated ESA612 standardizes test procedures, compares results to standards limits, and generates and stores reports for total digital data management.
Features Fluke ESA612
- Portable, ergonomic, lightweight and easy to use
- Large, easy-to-read display with adjustable contrast
- Human-factors-designed user interface
- Bail design for stand-up testing in field environments
- Five applied parts jacks and easy ECG snap connection with optional expander box
- ECG waveform tests and dual-lead measurements combine the functionality of a simulator, multimeter and electrical-safety analyzer in a single test tool
- Replaceable mains fuses keep the device in the field and out of the repair shop
- Internal memory for 100 test records
- USB connection for use with Ansur and Data Viewer software (for memory download to PC)
- Optional Ansur automation software standardizes test procedures, compares results to standards limits, generates/stores reports and provides total digital data management
- Rigorously tested for rugged field applications, with CE and CSA in addition to the Fluke-quality-design stamp of approval
More Information Fluke ESA612
- Automated: Pre-set templates allow you to test to the global standard of your choices at the push of a button. Let the unit guide you through the test steps quickly and accurately in three simple steps:
1. Select the test sequence corresponding to your standard of choice
2. Initiate the automated test sequence.
3. Store the results on-board or wirelessly export to your PC in seconds.
It is that easy. The automated test sequences are easily customized to suit individual testing requirements. - Portable: The ESA612 is a small, lightweight analyzer with an integrated handle to carry from place to place for on-the-go field service. It is designed for operation in tight spaces, and is easy to fit on any cart. A light, protective carrying case makes it easy to store and transport.
- Simple: A push-button interface allows quick access to highly-comprehensive functions and features, and an intuitive interface guides the user through tests. The ESA612 features a large display for clear indication of available functions, set-up criteria, device under test receptacle conditions, and test results. Data can be entered quickly with a plug–and-play keyboard, barcode scanner and/or on-board data entry interface.
Data archival is fast and simple with wireless communication or through a removable memory card with capacity for thousands of test results.
Specifications Fluke ESA612
Voltage |
|
Range (mains voltage) | 180 to 264 V ac rms |
Range (accessible voltage) | 0 V ac to 300 V ac rms |
Accuracy | ± (2 % of reading + 0.2 V) |
Voltage tests | Mains, accessible, and point-to-point |
Earth resistance |
|
Mode | Two terminal |
Test current | > 200 mA ac |
Range | 0 W to 2 W |
Accuracy | ± (2 % of reading + 0.015 W) |
Resistance tests | Earth resistance and point-to-point |
Equipment current |
|
Mode | AC rms |
Range | 0 A to 20 A |
Accuracy | ± 5 % of reading + (2 counts or 0.2 A, whichever is greater) |
Duty cycle | 15 A to 20 A, 5 min on/5 min off 10 A to 15 A, 7 min on/3 min off 0 A to 10 A continuous |
Leakage current |
|
Modes Modes are available in all leakage tests with the exception of MAP leakages that are available only in true-rms |
AC + DC (true-rms) AC only DC only |
Patient load selection (input impedance) |
AAMI ES1-1993 IEC 60601 |
Crest factor | = 3 |
Ranges | 0 µA to 199.9 µA 200 µA to 1999 µA 2 mA to 10 mA |
Frequency response/accuracy DC to 1 kHz 1 kHz to 100 kHz 100 kHz to 1 MHz |
± (1 % of reading + (1 µA or 1 LSD, whichever is greater)) ± (2 % of reading + (1 µA or 1 LSD, whichever is greater)) ± (5 % of reading + (1 µA or 1 LSD, whichever is greater)) |
Leakage tests | Ground wire (earth) Chassis (enclosure) Lead to ground (patient) Lead to lead (patient auxiliary) Lead isolation (mains on applied part) Direct equipment Direct applied part Alternative equipment Alternative applied part Point to point |
Mains on applied part test voltage | 100 % of mains |
Differential leakage |
|
Ranges | 10 µA to 199 µA 200 µA to 2000 µA 2 mA to 20 mA |
Accuracy | ± 10 % of reading + (2 counts or 20 µA, whichever is greater) |
Insulation resistance |
|
Ranges | 0.5 MW to 20 MW 20 MW to 100 MW |
Accuracy | ± (2 % of reading + 0.2 MW) ± (7.5 % of reading + 0.2 MW) |
Source test voltage | 500 V dc 250 V dc |
Insulation resistance tests | Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed accessible conductive part) and AP- NE (non-earthed accessible conductive part) |
ECG performance waveforms |
|
Accuracy | ± 2 % ± 5 % for amplitude of 2 Hz square wave only, fixed at 1 mV Lead II configuration |
Waveforms ECG complex (BPM) Ventricular fibrillation Square wave (50 % duty cycle) (Hz) Sine wave (Hz) Triangle wave (Hz) Pulse (63 ms pulse width) |
30, 60, 120, 180, and 2400.125 and 2 10, 40, 50, 60, and 100 2 30 BPM and 60 BPM |
Power ratings |
|
Mains voltage outlet | 230 V ac |
Mains voltage inlet power range | 180 to 264 V ac rms |
Maximum current | 16 A |
Hz | 50 or 60 |
Environmental specifications |
|
Operating temperature | 10 °C to 40 °C (50 °F to 104 °F) |
Storage temperature | -20 °C to 60 °C (-4 °F to 140 °F) |
Operating humidity | 10 % to 90 % non-condensing |
Altitude | 230 V ac mains supply voltage up to 2,000 m |
Package Include :
- Operator’s Manual (multilingual CD)
- Getting Started Guide (printed, multilingual)
- Ansur ESA612 Plugin, CD with demo version
- Data Transfer Cable
- Empty Post Adapter
- Banana Jack – ECG Adapter
- Carrying Bag
- The power cable supplied is one piece and is country specific depending on the model number.
Additional information
Lead Time | 5-7 days |
---|---|